Assembly Replacement Procedures

Radiated Filter Board Assembly

Radiated Filter Board Assembly

Removal

1. Remove the instrument outer case. Refer to the
procedure.
2. Remove the top brace. Refer to the
3. Refer to
Figure
7-21. Remove W9, W10, and W12.
4. The Radiated Filter Board assembly (2) can be removed from the chassis by
pulling straight up using the extractors.

Replacement

1. Slide the Radiated Filter Board assembly (2) into the slot as shown in
7-21, using the extractors to secure the board to the connectors on the
motherboard assembly.
2. Replace W9, W10, and W12. Torque to 10 inch-pounds.
3. Replace the top brace. Refer to the
4. Replace the instrument outer case. Refer to the
replacement procedure.
194
Instrument Outer Case
Top Brace
removal procedure.
Top Brace
replacement procedure.
Instrument Outer Case
removal
Figure
Chapter 7