Replacement

1. Slide the power supply assembly into the slot at the rear of the instrument and
push on the assembly to mate the connectors to the midplane assembly.
2. Refer to
Figure
7-11. Replace the screws (2) and (3) through the power
supply bracket (1) and into the power supply (4) and chassis. Torque to 9
inch-pounds.
3. Replace the top brace. Refer to the
4. Replace the rear panel. Refer to the
5. Replace the processor assembly. Refer to the
procedure.
6. Replace the instrument outer case. Refer to the
replacement procedure.
Chapter 7
Assembly Replacement Procedures
Power Supply Assembly
Top Brace
replacement procedure.
Rear Panel
replacement procedure.
Processor Assembly
Instrument Outer Case
replacement
177