Troubleshooting
Instrument Boot Up Process
7. If the recovery system is not selected the Windows XP Professional operating
If a recovery was selected follow the on-screen instructions and perform a system
NOTE
recovery. Additional information about performing a system recovery can be found
in the
8. Agilent Technologies logo is displayed in white font on a blue background
9. By default, the initialization process of the N9039A RF Preselector application
Once the N9039A RF Preselector application is fully initialized and aligned, the
display should resemble
AC power through the loading of the N9039A RF Preselector application software.
Figure 2-1
Instrument Display at Power-up
36
system will begin to boot up. This will take ~20-30 seconds.
"A5 Disk Drive Assembly"
while Windows finishes loading user preferences. If this does not occur refer to
the
"Instrument Cannot Completely Load or Run the Operating System"
section in this chapter.
begins loading. If any of the initializing processes did not complete, refer to the
"Initializations Did Not Complete"
Figure
section in this chapter.
section in this chapter.
2-1. This completes the boot process from initial