Using The Agilent Recovery System - Agilent Technologies N9039A Service Manual

Rf preselector
Hide thumbs Also See for N9039A:
Table of Contents

Advertisement

Troubleshooting
A5 Disk Drive Assembly
• Additional software that was installed after the instrument was shipped from
• Any updates that were made to the N9039A RF Preselector application

Using the Agilent Recovery System

To run the Agilent Recovery System on the instrument perform the following
procedure:
1. Make sure the instrument is turned off.
2. Turn on the instrument.
3. After the "Agilent Technologies" splash screen is displayed you will be
4. Press the down arrow key to move the highlight to "Agilent Recovery System"
5. When the Agilent Recovery System has booted, follow the on-screen
6. After exiting the Agilent Recovery System, the instrument will reboot a few
7. Once the Agilent Recovery System has completed, update the N9039A RF
The latest version of the N9039A RF Preselector application software needs to be
NOTE
installed after an A5 Disk Drive is replaced, even if the replacement drive already
has the latest version on, as this is the only way to update the FPGA code on the
A3 Digital I/O board assembly, which might have had an older version prior to the
drive replacement.
60
the factory. After a recovery, that software will need to be re-installed by the
end user.
software.
presented with a screen as seen in
make a selection as outlined in step 4.
and press the
key.
Enter
instructions to recover the C drive.
times, during which time you will be asked to accept the End User License
Agreement.
Preselector application software to the latest version by downloading it from:
www.agilent.com/find/N9039A_software
Figure
2-14. You will have 5 seconds to
Chapter 2