Figure 7-3
Instrument Outer Case Removal
Replacement
1. Disconnect the instrument from ac power.
2. Slide the instrument cover back onto the deck from the rear. The seam on the
3. Replace the four rear feet to the rear of the instrument. Torque the rear feet
4. Replace the handle straps on both sides of the instrument. Torque the handle
Chapter 7
cover should be on the bottom. Be sure the cover seats into the gasket groove in
the Front Frame Assembly.
screws to 21 inch pounds.
strap screws to 21 inch pounds.
Assembly Replacement Procedures
Instrument Outer Case
167