Agilent Technologies N9039A Service Manual page 131

Rf preselector
Hide thumbs Also See for N9039A:
Table of Contents

Advertisement

Radiated Path
Selects the radiated filter path. Once selected the service technician can then
select one of the 9 different radiated filter paths.
Band 0
Selects the 28 MHz to 55 MHz filter path
Band 1
Selects the 55 MHz to 100 MHz filter path
Band 2
Selects the 100 MHz to 155 MHz filter path
Band 3
Selects the 155 MHz to 235 MHz filter path
Band 4
Selects the 235 MHz to 350 MHz filter path
Band 5
Selects the 350 MHz to 470 MHz filter path
Band 6
Selects the 470 MHz to 620 MHz filter path
Band 7
Selects the 620 MHz to 820 MHz filter path
Band 8
Selects the 820 MHz to 1 GHz filter path
Through Path
Selects the internal through path. Once selected the service technician can then
select one of the 3 different internal through paths. These are different than the
Bypass Path in that the Through Bands are internal to the filter boards, unlike
the Bypass Path which does not route the input signal to the filter boards at all.
Band 0
Selects the 9 kHz to 2 MHz through path
Band 1
Selects the 2 MHz to 30 MHz through path
Band 2
Selects the 28 MHz to 1 GHz through path
Chapter 5
Service and Diagnostics Menus
Service Key Descriptions
131

Advertisement

Table of Contents
loading

Table of Contents