Trigger Functions; Alarm Functions - GRAPHTEC GL2000 User Manual

Midi logger hv
Hide thumbs Also See for GL2000:
Table of Contents

Advertisement

CHAPTER 4 Specification

Trigger Functions

Item
Repeat Trigger
Repeat interval
Trigger types
Trigger conditions
Trigger judgment modes
Combination of channel

Alarm Functions

Item
Alarm judgment modes
Alarm detection cycle
Alarm generation hold function
Combination of channel
4-8
Off, On
Start to start interval, stop to start interval
1 second - 9999 hours 59 minutes 59 seconds
Start: Data capture starts when a trigger is generated.
Stop: Data capture stops when a trigger is generated.
Start: Off, level, alarm, external input, specified time, specified day of the week, certain
time, hourly
Stop: Off, level, alarm, external input, specified time, specified day of the week, certain time
* Level can be set to each CH.
Analog judgment: H/Rising, L/Falling, Window In, Window Out
Logic judgment: H, L (Pattern comparison)
Pulse judgment: H/Rising, L/Falling, Window In, Window Out
Level OR, Level AND, Edge OR, Edge AND
Analog judgment: H, L, Window In, Window Out
Logic judgment: H, L
Pulse judgment: H, L, Window In, Window Out
Linked to the sampling interval (Fastest 1 μs)
ON, OFF Setting available
Level OR fixed
Description
Description

Advertisement

Table of Contents
loading

Table of Contents