4.2 Function Specifications

Function Specifications
Item
Display screen
EU (scaling function)
Review function
Calculation
Search functions
Annotation input function

Trigger Functions

Item
Timer mode
Repeat trigger
Trigger types
Trigger settings
Level judgment modes
Channel combinations
Waveform screen + Digital screen
Expanded Waveform screen
Digital screen
X-Y display
* Even during data capture, you can open menus (to check whether setting is possible).
* Screens can be key-toggled.
4 points can be set for each channel.
Data replay during data capture
Types of statistical calculation : Average, Max, Min, Peak, RMS
Number of operations
Method
Function
: Search the captured data for the required number of points
Search type : Search of channels by levels
Search by logic pulses + combinations
Search by alarm generation
Function
: A comment can be input for each channel.
Supported characters : Alphanumeric and kana characters
Number of characters : 11 (Up to 8 characters are displayed)
Off, Date and Time, Every Day Cycle, Every Hour Cycle
Off, On
Start
: Data capture starts when a trigger is generated.
Stop
: Data capture stops when a trigger is generated.
Start
: Off, Level, External
Stop
: Off, Level, External, Time
* A level can be set for each channel.
Analog : ↑ H, ↓ L, Window IN, Window Out (Tolerance ±1%)
Logic : ↑ H, ↓ L
Pulse : ↑ H, ↓ L, Window IN, Window Out
Level OR, Level AND, Edge OR, Edge AND
Description
: Maximum of 2 can be set simultaneously.
: Data between cursors specified (during data replay)
Description
Specifications
4-5