GRAPHTEC GL2000 User Manual page 113

Midi logger hv
Hide thumbs Also See for GL2000:
Table of Contents

Advertisement

CHAPTER 3 Settings and Measurement
(2)-3 Capturing destination
Set the capturing destination of the measured data.
Setting
Internal RAM
Internal memory/
SD card/
USB memory
Measured data
3-72
The measured data is captured to the internal RAM. The data is lost when you turn off the power.
When Overwrite is set to On, the data is overwritten if the next capturing is performed.
- "7. File menu".)
Additionally, When Auto Save is set to On, the data is saved automatically.
You can use all the sampling intervals. Select and capture at speed faster than 500 µs.
Also, when the waveform of the captured data is checked to save only the necessary data, the
memory can be saved for that instance.
The measured data is captured to the internal memory/SD card/USB memory. The captured data is
saved even when the power is turned off. This cannot be selected when the sampling interval is 1 to
500 µs. In this case, set the sampling interval to a value slower than 1 ms.
You can also use large capacity SD card (*1)/USB memory (*2) (*3). When the capacity of the
more intervals or you measure for a longer period.
(*1) SD card and SDHC card are available. SDXC memory card is not available.
(*2) FAT32/FAT16 format is available. NTFS, exFAT format is not available.
authentication and the USB memory without shell (metal part) in the connector part.
For check the latest information and support information, please check with the following URL.
http://www.graphteccorp.com/
Data lost after
Data retained
power-off
after power-off
Saving data
or Auto Save
SD card memory or
Internal RAM
USB memory
Description

Advertisement

Table of Contents
loading

Table of Contents