CHAPTER 3 Settings and Measurement
(5)-3 TCP/IP advanced settings
Setting
(1) IPAddr Auto Acq.
(2) IP Address
(3) Sub-net mask
(4) Port number
(5) Gateway
(6) DNS address
(7) Keep Alive
(8) Apply Settings
If the Automatic IP Address Acquisition fails (see the figure below), the IP address is not set properly.
Disable the Auto IP Address Acquisition and make the settings one by one .
After you have changed the TCP-IP settings, turn off and turn on the power, or execute Reflect Settings (The
connection will force disconnect) .
When you want to use the IP address automatic acquisition function, DHCP server must be operating
separately within the searchable network .
< Auto IP Address Acquisition is successful >
3-108
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
Set the IP address automatic acquisition.
On, Off
Set the IP address.
0 to 255 . 0 to 255 . 0 to 255 . 0 to 255
* When the IP address is set automatically, the automatic set IP address is displayed.
Set the subnet mask.
0 to 255 . 0 to 255 . 0 to 255 . 0 to 255
* When the IP address is set automatically, the automatic set sub-net mask is displayed.
Set the port number of the control command server.
1024 to 65535
* This is the port number specified when connecting to our application (GL980-2000_APS).
Set the gateway.
0 to 255.0 to 255.0 to 255.0 to 255
※ When IP address is set automatically, the automatic set gateway address is displayed.
Set the address of the DNS server.
0 to 255 . 0 to 255 . 0 to 255 . 0 to 255
* When the IP address is set automatically, the automatic set DNS server address is
displayed.
Set forced disconnection time when no communication is detected by the control command
server.
Off, 10 seconds, 30 seconds, 1 minute, 10 minutes, 30 minutes, 1 hour
If no communication state continues for specified time or more, the socket connection is
disconnected. You should generate a communication within the set time. When using the
supplied application software, note that it will be the non-communication state while the
captured data is being replaying.
(This function is available only for the control command server, and does not affect the Web
server function and FTP server function.)
Confirm TCP/IP setting and restart TCP/IP processing.
< Auto IP Address Acquisition fails >
Description

Advertisement

Table of Contents
loading

Table of Contents