(5)-4 FTP/WEB server settings
Set the FTP server/WEB server as this device.
Selection item
FTP Client
(1) FTP Server
(2) User Name
(3) Password
(4) Port Number
(5) PASV Mode
(6) FTP Server
Connection
Test
FTP server
(7) Port number
WEB server
(8) WAB number
(5)-5 Mail settings
Enter the settings for emails on the device.
When mail send setting is complete various notification setting (Alarm, low battery, drive low capacity (only
during capturing)) is sent. Also, when periodic notification is set, it is sent according to the set time.
(1) Mail send settings
Setting
Outgoing mail
Destination
settings
Settings
Notification
settings
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
Enter the domain name or IP address of the FTP server. (Up to 127 characters)
Enter the user name of the FTP account. (Up to 31 characters)
Enter the password of the FTP account. (Up to 31 characters)
Enter the port number of a port to be used for FTP. It is normally 21. (0 to 65535)
Make the passive mode setting.
ON : Should be set for communication with an external FTP server in a firewall environment.
OFF : Should be set for communication with an FTP server in a normal network environment.
Press right
key to execute (Performs connection test to the FTP server.)
When the connection test is performed, a message is displayed.If connection cannot be
established, check the settings and perform the connection test again.
* If the connection test is passed, the following message is displayed.
Set port number of the FTP server installed in this device. (0 to 65535)
Set port number of the WEB server installed in this device. (0 to 65535)
(1)
(2)
TO
Set the destination (address) of the E-mail.
Enter the string of up to 63 characters.
CC1 to CC3
Set the destination of the E-mail (broadcast 1 to 3).
Enter the string of up to 63 characters.
Subject
Set the subject of the E-mail.
Enter the string of up to 63 characters.
Alarm
Set the E-mail send when an alarm occurs.
Off, On
Low battery
Set the E-mail send when the internal battery lowers.
Off, On
Drive free space
Set the E-mail send when the drive free capacity during recording is
low.
Off, On
Periodic
Set the periodic notification E-mail send.
notifications
Off, 1 hour, 2 hours, 3 hours, 6 hours, 12 hours, Specified time
* You can set the specified time to the hours : minutes : seconds.