GRAPHTEC GL2000 User Manual page 130

Midi logger hv
Hide thumbs Also See for GL2000:
Table of Contents

Advertisement

CHAPTER 3 Settings and Measurement
Position (XY position)
Set the XY position.
This setting is used for XY display only.
This differs from the position setting for Y-T waveform display (The position setting that can be operated at Y-T
waveform + Digital display.)..
For XY position setting, set the drawing reference position to 0% for the lower left and 100% for the upper
right of the XY graph.
100%
50%
0%
0%
50%
100%
The measurement range displayed on the graph by the XY position setting is as follows.
-10DIV
-5DIV
Reference point
5DIV
10DIV
Position 100%
Position 70%
Position 50%
Position 10%
Position 0%
As shown in the figure above, the display range moves 10DIV area in parallel through the XY position
setting.
XY position setting is set only for XY graph.
Position reference point differs between range mode and span mode.
• Range mode
-FS
Reference point
+FS
0V, 0Vrms
The position reference point in Range mode is 0V or 0Vrms.
3-89

Advertisement

Table of Contents
loading

Table of Contents