Sweep Time; Output Power - Keysight Technologies FieldFox N9912A104 User Manual

Table of Contents

Advertisement

NA (Network Analyzer) Mode
NA Mode Settings

Sweep Time

The fastest possible sweep time is always used as the default setting. Use the
Min Swp Time setting to slow the sweep time when measuring long lengths of
cable. Learn more at the FieldFox Supplemental Online Help:
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
The actual sweep time is shown on the FieldFox screen. See the Screen Tour
on
"Screen Tour" on page
is higher than the actual sweep time. The increase will not be exactly the
amount that you enter, as the actual sweep time is the composite of many
factors.
Measurement speed specifications do NOT apply in Temperature Control
Mode. Learn more in
How to set Sweep Time
— Press Sweep 3.
— Then
— Enter a value using the numeric keypad.
— Press a multiplier key. Learn about multiplier abbreviations in
Abbreviations" on page

Output Power

Set the power level out of the FieldFox to High, Low, or manually set power
level to a value between High and Low.
Generally, the high power setting is used when measuring passive, high-loss
devices to place the signal farther from the noise floor. However, for devices
that are sensitive to high power levels such as amplifiers, use the Low power
setting.
For best measurement accuracy, use the Manual power setting at -15 dBm.
After calibration, the power level can be decreased for amplifiers, or increased
for higher dynamic range.
Power Level settings in this mode will NOT change Power Level settings in
other modes. To help prevent damage to your DUT, use caution when
changing modes with your DUT connected to the FieldFox test ports.
How to set Output Power
— Press Meas Setup 4.
— Then
5- 14
2-12. To increase the sweep time, enter a value that
"Temperature Control Mode" on page 2-5
Min Swp
Time.
2-14.
Output
Power.
.
Keysight N9912-90001 User's Guide