Keysight Technologies FieldFox N9912A104 User Manual page 63

Table of Contents

Advertisement

DTF (Distance to Fault) Measurements
DTF Measurement Settings
A true fault response will not move in distance. That is, if a true fault is present
at 10.3 meters, changing the stop distance from 15 m to 20 m will not move
the fault; the fault will remain at 10.3 meters. However, an alias response will
appear to move.
An un-terminated cable (with NO perfect load at the end) will show faults that
appear to be beyond the end of the cable. These are NOT alias faults. These
faults appear as the signal reflects off the open at the end of the cable and
travels back down the cable toward the connection at the FieldFox.
Re-reflections are measured at the FieldFox as mirror images of the original
faults. The largest fault is the open end of the cable. To avoid confusion, set
the Stop distance shortly after that fault.
Keysight N9912-90001 User's Guide
4- 27

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fieldfox n9912aFieldfox n9912a106

Table of Contents