File Management
Saving and Recalling Files
In the graphic above, the current filename is File_0. The cursor is on the
custom
Type is "Trace + State" (Instrument State).
How to move the cursor
— Use the rotary knob to move cursor on the same row. The cursor will change
rows when it comes to a row end.
— Use the
When the current filename is complete
— Press
exists, a warning is displayed.
— Press
— If the filename already exists, you are prompted to choose from the
following:

Recall Files

The FieldFox can recall State files, Trace + State files, and image files.
State Files
ALL FieldFox instrument settings for ALL Modes are saved and recalled in a
*.sta file. If error correction is ON in CAT or NA Mode, then calibration data is
also saved and recalled.
If error correction is OFF (CalRdy is active), calibration data is NOT saved in the
*.sta file. When a Trace + State file is recalled on a FieldFox other than the one
it was saved on, differences in the CalRdy calibration data may result in
differences between the displayed traces.
Image Files
The FieldFox can open and display *.png, *.jpg, and *.jpeg images.
How to Recall a File
Before recalling a file, first set File Type to State or Trace + State, and also
Select Device from which the file will be recalled.
— Press Save/Recall 9.
— Then Recall.
17- 4
DTF_VSWR
keyword. The file will be saved to USB (flash drive). File
arrows to move the cursor up and down the rows.
Done
to save the file to the specified filename. If the filename already
Cancel
to close the labeler without saving a file.
— Press
Yes
to overwrite the existing file.
— Press
Cancel
to return to edit the filename.