Display Resolution; If Bandwidth; Output Power - Keysight Technologies FieldFox N9912A104 User Manual

Table of Contents

Advertisement

VVM (Vector Voltmeter) Mode - Option 308
VVM Mode Settings

Display Resolution

You can display either one digit or two digits after the decimal point for both
magnitude and phase readings.
This setting survives a Preset and Mode Preset.
How to set Resolution
— Press Scale/Amptd .
— Then Resolution.
— Then choose:

IF Bandwidth

Reducing the IF receiver bandwidth reduces the effect of random noise on a
measurement. However, narrower IF bandwidths cause longer measurement
times.
For highest accuracy, Zero AFTER changing the IF Bandwidth. Learn about
"Zeroing" on page
How to set IF BW
— Press BW 2.
— Then
— Then choose from the following:
300 Hz | 1 kHz | 3 kHz | 10 kHz | 30 kHz

Output Power

Set the power level out of the FieldFox to High, Low, or manually set power
level to a value between High and Low.
Generally, the high power setting is used when measuring passive, high-loss
devices to place the signal farther from the noise floor. However, for devices
that are sensitive to high power levels such as amplifiers, use the Low power
setting.