Calibration for NA, CAT, and VVM Modes
Why and When to Calibrate
2. The Frequency Range, Resolution, Power (Hi/Low), and IF BW (in NA Mode)
settings of the two calibrations are identical.
When both conditions occur, you can switch between the 1-port and the
2-port measurements, and both will remain fully calibrated.
In a DTF measurement, the frequency range over which the calibration is made
can be viewed using the calculated Start and Stop Frequency on the Settings
page. Learn how to see this in
"Calculated DTF values" on page
4-26.

Save the Calibration

After performing any type of calibration, you can save the FieldFox settings
along with the calibration into a STATE (*.sta) file. These settings and
calibration can then be recalled as necessary. To learn how, see
"Saving and
Recalling Files" on page
17-2.
Keysight N9912-90001 User's Guide
7- 9