Gating; How To Make Time Domain Gating Settings; Start, Stop, Center, And Span Gate Times; Gating Type - Keysight Technologies FieldFox N9912A104 User Manual

Table of Contents

Advertisement

Time Domain – Option 010

Gating

Gating
Perhaps the most beneficial feature of Time Domain transform is the Gating
function. When viewing the Time Domain response of a device, the gating
function can be used to "virtually" remove undesired responses. You can then
simultaneously view a frequency domain trace as if the undesired response did
not exist. This allows you to characterize devices without the effects of external
devices such as connectors or adapters.
Gating is best performed while viewing a Time Domain trace. Then, either
disable Transform, or separately view a frequency domain trace to see the
trace with the gated effects removed.
When a discontinuity in a test device reflects energy, that energy will not
reach subsequent discontinuities. This can "MASK", or hide, the true
response which would have occurred if the previous discontinuity were not
present. The Gating feature does NOT compensate for this.

How to make Time Domain Gating settings

— Press Measure 4 .
— Then Transform.
— Then Gating.
— Then choose from:

Start, Stop, Center, and Span Gate Times

These settings specify the time in the trace to be gated (in or out). The gate
times can be specified using either Start and Stop or Center and Span.
— Press
— A settings table appears that allows you to navigate and select the

Gating Type

This setting defines the type of filtering that will be performed for the gating
function. The gate start and stop flags on the display point toward the part of
the trace you want to keep.
Choose from the following:
6- 10
On
Gating is being performed.
Off
Gating is NOT being performed.
Gating Start
Stop.
following settings. These settings are made exactly like those made on the
Transform Settings table. Learn how in
page
6-4.
Bandpass
KEEPS the responses within the Gating Start and Stop times.
"Transform Settings Table" on
Keysight N9912-90001 User's Guide