Connector Care Review
Handling and Storage
— Keep connections clean
— Extend sleeve or connector nut
— Use plastic end-caps during storage
Visual Inspection
— Inspect all connectors carefully
— Look for metal particles, scratches, and dents
Connector Cleaning
— Try compressed air first
— use isopropyl alcohol
— Clean connector threads
Gaging Connectors
— Clean and zero the gage before use
— Use the correct gage type
— Use correct end of calibration block
— Gage all connectors before first use
Making Connectors
— Align connectors carefully
— Make preliminary connection contact lightly
— Turn only the connector nut
— Use a torque wrench for final connection
a. Cleaning connectors with alcohol shall only be done with the DC power cord removed, in a well-venti-
lated area, and with the power to the FieldFox turned OFF. Allow all residual alcohol moisture to evapo-
rate, and the fumes to dissipate prior to energizing the instrument.
A- 2
Do
Do
Do
a
Do
Do
Do Not
— Touch mating-plane surfaces
— Set connectors contact-end down
— Store connectors or adapters loose
Do Not
— Use a damaged connector—ever
Do Not
— Use any abrasives
— Get liquid into plastic support beads
Do Not
— Use an out-of-specification connector
Do Not
— Apply bending force to connection
— Over Tighten preliminary connection
— Twist or screw any connection
— tighten past torque wrench "break" point