Keysight Technologies FieldFox N9912A104 User Manual page 134

Table of Contents

Advertisement

SA (Spectrum Analyzer) Mode
SA Mode Settings
Trigger Delay
After a valid External or Video trigger signal is received, the sweep begins after
the specified Trigger Delay time. Only positive values are allowed.
To see the rising edge of a repetitive signal which is triggered on that edge, use
trigger delay to clearly view the subsequent signal one period later.
— Press Sweep 3.
— Then
— Then
— Enter a value using the numeric keypad, the
knob.
— Then select a multiplier key or press Enter. Learn more in
Abbreviations" on page
Trigger Level
Used with Video triggering, an incoming signal with this amplitude or higher
(for positive slope) will initiate a sweep. The Units depend on the Scale Type
setting: (Log or Linear). Learn more in
The Video Trigger Level is a time domain (Zero Span) signal level comparison.
Therefore, the sweep will trigger close to the displayed level in Zero Span
measurements. In frequency span measurements, processing can cause
broadband signal energy to display at lower power levels than the originating
time domain signal. Therefore, you may need to set the trigger level higher
than the displayed level.
— Press Sweep 3.
— Then
— Then
— Enter a value using the numeric keypad, the
knob.
— Then select a multiplier key or press Enter. Learn more in
Abbreviations" on page
Auto Trigger Time
If an External or Video trigger signal is not received before the specified Auto
Trig Time, a sweep will occur automatically.
Enter 0 to set Auto Trigger OFF. When Auto Trigger is OFF, the FieldFox does
NOT sweep unless a valid External or Video trigger signal is received.
— Press Sweep 3.
— Then
8- 26
Trigger
Settings.
Trig
Delay.
2-14.
Trigger
Settings.
Trig
Level.
2-14.
Trigger
Settings.
arrows, or the rotary
"Scale and Units" on page
arrows, or the rotary
Keysight N9912-90001 User's Guide
"Multiplier
8-8.
"Multiplier

Advertisement