SA (Spectrum Analyzer) Mode
SA Mode Settings
— Then
— Enter an Auto Trig Time using the numeric keypad.
— Then select a multiplier key. See

Single / Continuous / Restart

The following behavior is unique to SA Mode.
This setting determines whether the FieldFox measures continuously or only
once each time the
Single or to conserve battery power or to allow you to save or analyze a
specific trace.
How to set Single or Continuous
— Press Sweep 3.
— Then choose one of the following:

Points

Points is the number of measured frequencies the X-axis. The higher number of
data points, the better the ability to resolve closely spaced signals and the
slower the sweep speed.
How to set number of points
— Press Sweep 3.
— Then
Keysight N9912-90001 User's Guide
Auto Trigger [current
Single
or Run / Hold +/- button is pressed. Use Hold /
Continuous
——ON
Causes the SA to make continuous sweeps. This is the
typical setting for making measurements. Each Run / Hold +/-
press causes the SA to toggle between continuous sweep and
Hold.
——OFF
Causes the SA to Hold. Each Run / Hold +/- or
press causes the SA to make ONE sweep, then Hold. The Hold
annotation changes to an arrow --> while the sweep occurs.
Single
Automatically sets Continuous OFF and causes the SA to
make ONE sweep, then Hold.
Restart
Resets the Average counter on 1. When in Hold, causes the
SA to make ONE sweep, then Hold. The Average counter has is used
with MaxHold and MinHold Trace States. Learn more in
Count" on page
8-29.
Points [current
setting].
setting].
"Multiplier Abbreviations" on page
2-14.
Single
"Average
8- 27