4.6

Use the BOND-PRIME Cleaning Kit

If required:
4.2 Refill the DI Water Container
l
4.3 Refill the Alcohol Container
l
4.4 Refill the Lot Tracked Bulk Containers
l
4.5 Empty the Waste Containers
l
The BOND-PRIME Cleaning Kit cannot be preemptively scheduled.
Before you start this procedure, ensure you are logged in to the processing module. Refer to
in and log
out.
BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
A message is displayed in the Action Queue when it is time to
run the BOND-PRIME Cleaning Kit. ARC Modules must be
cleaned when the usage count is between 17 and 23.
You must register the BOND-PRIME Cleaning Kit on the BOND
System Controller (refer to BOND System User Manual).
Some ARC Modules are disabled until cleaning is completed.
Ensure there are no slides in the Preload
Drawer and Unload Drawer when starting a
clean.
1.
Retrieve the Reagent Tray containing the BOND-PRIME
Cleaning Kit.
Chapter  4 Cleaning and maintenance
2.1 Log
123