2.5

Preload slides

WARNING: You must wear the minimum required Personal Protective Equipment (PPE) before you
use reagents or operate the processing module. Refer to
Before you start this procedure, ensure you are logged in to the processing module. Refer to
in and log
out.
BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
 General
Make sure the slide labels are correctly affixed and positioned
wholly on the slide.
Do not stack more than two labels on a slide.
Slides must be free of:
excess moisture and debris, for example dust, wax and
l
glass chips.
sticky residue from removed/re-affixed labels.
l
1.
Bake slides at 60 °C (140 °F) for 60 minutes to improve
tissue adhesion.
The Preload Drawer is the left-hand drawer.