4.16 Clean the Locked Bulk Reagent Containers
WARNING: You must wear the minimum required PPE before you maintain the processing module.
Refer to
General
WARNING: Take care not to collide with the Bulk Container Drawers while they are open.
Before you start this procedure, ensure you are logged in to the processing module. Refer to
in and log
out.
BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
cautions.
1.
Tap Status.
2.
Check the Status screen to ensure there are:
Refer to
no slides currently being processed (Processing)
l
no slides in the Preload Drawer (Loaded) and
l
Unload Drawer (Complete).