5.3
Manually retrieve slides from the processing
module
The processing module might occasionally indicate that slide processing cannot continue and you need to manually
retrieve slides. You can retrieve slides from the following locations:
Preload Drawer—refer to
l
Unload Drawer—refer to
l
Work Surface—refer to
l
ARC Modules—refer to
l
There are a number of reasons why slides might be dropped. These include:
the Suction Cup on the High-Speed Robot needs cleaning or replacement. Refer to
l
or
4.11 Replace the Suction Cup
the label has been incorrectly placed on the slide, or there are more than two labels on the slide. Refer to
l
label specifications
there is tissue, residue, or liquid on the slide label area.
l
there is an issue with the Vacuum System
l
the Preload Drawer or Unload Drawer has been replaced and is out of alignment.
l
The Action Queue will display a message indicating the cause of the issue and the action required to rectify it.
If slide transfer continuously fails, contact customer support.
5.3.1
Manually retrieve slides from Preload and Unload
Drawers
WARNING: You must wear the minimum required PPE before you maintain the processing module.
Refer to
 General
If prompted by a message in the Action Queue to remove slides from the Preload or Unload Drawer, follow this
procedure.
BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
5.3.1 Manually retrieve slides from Preload and Unload Drawers
5.3.1 Manually retrieve slides from Preload and Unload Drawers
5.3.2 Manually retrieve slides from the Work Surface
5.3.3 Manually retrieve slides from ARC Modules
cautions.
Chapter  5 Troubleshooting
4.10 Clean the Suction Cup
 Slide
190