4.1.3
Cleaning and maintenance checklist
DAILY/AS REQUIRED
Examine Action Queue and Alert Banner
Refill/empty bulk containers as required
BOND-PRIME Wash Solution Concentrate lot number
ER1 lot number
ER2 lot number
Dewax Solution lot number
WEEKLY
Wipe internal surface of ARC Modules
Wipe down Reagent Platform and ARC Bank
Surfaces
Clean Suction Cup
TWO-MONTHLY
Replace Suction Cup (3400 slides)
Clean Slide Drawer Inserts, Waste Drain and Sumps,
and Unload Drawer Pickup Filter*
Clean Wash/Prime Stations*
Clean Sump Tray*
Clean external surfaces
Clean Barcode Scanner (on Hood)
FOUR-MONTHLY/5000 slides
Use the BOND-PRIME ARC Refresh Kit
SIX-MONTHLY
Clean all bulk and waste containers
Clean hand-held barcode scanner
(on BOND System Controller)
NOTIFICATION
Use the BOND-PRIME Cleaning Kit
For week starting ___________________to ____________________ For month of _____________________
BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
Mon
Tues
Wed
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
Chapter 4 Cleaning and maintenance
Thurs
Fri
Sat
*
*
*
* If it is necessary, do these tasks more
frequently.
Sun
*