4.11 Replace the Suction Cup
Replace the Suction Cup every 3400 slides or 2 months, whichever comes first.
To help prevent dropping the Suction Cup into the processing module, carefully move the High-Speed Robot over the
Reagent Platform.
WARNING: You must wear the minimum required PPE before you maintain the processing module.
Refer to
General
Before you start this procedure, ensure you are logged in to the processing module. Refer to
in and log
out.
BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
cautions.
1.
Tap Status.
2.
Check the Status screen to ensure there are:
Refer to
no slides currently being processed (Processing)
l
no slides in the Preload Drawer (Loaded) and
l
Unload Drawer (Complete).
2.3 Status
Screen.
Chapter 4 Cleaning and maintenance
2.1 Log
141