Unexpected events during processing

If you see an Action Queue Item (AQI) regarding an unexpected event that occurred during processing, you must take
extra steps to confirm that the flagged slide is suitable for diagnostic use.
This notification does not necessarily indicate that staining was in any way unsatisfactory, however the system
operator or laboratory supervisor must:
Examine the Slide history screen on the BOND System Controller (refer to the BOND System User Manual).
1
Look for a slide with the status Done (events noted) that has the same Slide ID as the flagged slide removed
2
from the BOND-PRIME Processing Module.
Select this slide then click the Run events button to generate the Run Events Report.
3
Any events that caused a notification are displayed in Bold text. The system operator or laboratory supervisor
should carefully consider the notification events listed, as these provide important details about the nature of
the slide notification events.