1.12.4 ARC (Active Reagent Control) Modules

Twenty-four ARC Modules mounted on two ARC Banks hold slides during staining. Collectively, they are known as the
ARC Array. Reagents are dispensed by the Bulk Reagent Probes and the ARC Probe on the Robot Head. ARC Modules
are cleaned using DI Water and BOND Wash Working Solution dispensed by the Wash Robots.
Waste from the ARC Modules is directed to the Hazardous Waste Reservoir.
Legend
1 ARC Module Latch
2 ARC Module Cover
If the Action Queue (see
Action Queue and Alert Banner on page
it has a Covertile.
If an ARC Module is leaking, check the:
condition of the Covertile and Covertile seal, and if necessary, replace the Covertile.
l
ARC Module for debris, and if necessary, clean
l
BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
3 ARC Module Lid Assembly
4 ARC Covertile:
71) indicates that an ARC Module is faulty, check that
(Wipe internal surface of ARC Modules on page
Chapter  1 BOND-PRIME hardware
129).
47