4.18 Clean the Sump Tray

WARNING: You must wear the minimum required PPE before you maintain the processing module.
Refer to
 General
BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
cautions.
1.
Locate the Sump Tray.
2.
Stand to the side of the Sump Tray and remove it using two
hands to avoid spillage. If required, follow the spill
management procedures that apply at the laboratory site.
3.
Soak up the excess waste reagent.
a.
Absorb the bulk of the waste reagent using paper
towels.
b.
Dispose of the paper towels in accordance with
laboratory procedures. Always treat the waste from the
Sump Tray as hazardous.
If there is an excessive amount of waste in
the Sump Tray, contact Leica Biosystems
Support.
Chapter  4 Cleaning and maintenance
179