Slides not accepted
On rare occasions, the processing module may fail to evaluate and accept a loaded slide.
When this happens, the slide remains at the bottom of the segment for an extended period.
To rectify this situation, you should:
Remove the slide from the Preload Drawer, then close the drawer.
1
Wait for the processing module to scan the drawer and update the Status screen.
2
Reload the slide into the Preload Drawer. The processing module should now be able to identify the slide and
3
accept it for processing.
2.3.2

Slides Processing Segment

To alternately show the Processing and Start now slide count, tap the displayed value.
When slides are transferred to the ARC Modules for processing, they appear at the right of the segment, earliest first.
There is no relationship between slide positions in the Slides Processing Segment and in the ARC Modules.
Also refer to
2.3.4 View more slide
Status Screen updates
The processing module may not always prioritize loading slides. This is because the processing module automatically
performs a basic clean of each ARC Module in between processing slides.