Clean the Waste Drains and Sump

BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
12.
Moisten a clean lint-free cloth with 70% ethanol solution.
13.
Wipe the Waste Drains with the lint-free cloth. The Waste
Drains are accessible when the Slide Drawer Inserts have
been removed from the opened drawers.
14.
Wipe the Sumps with the lint-free cloth. The Sumps are
located under and behind the Preload and Unload Drawers.
They are accessible via the Work Surface after fully
opening the drawers.