4.14 Use the BOND-PRIME ARC Refresh Kit

The BOND-PRIME ARC Refresh Kit contains replacement:
Covertiles
l
Mixing Well Plate.
l
Use the ARC Refresh Kit every 5000 slides or 4 months, whichever comes first. Each IHC single stain counts as one
usage for Covertile life purposes. Each ISH hybridization counts as two usages. The maximum usage per slide is
capped at two. The table below indicates the equivalent slide count per slide type.
WARNING: You must wear the minimum required PPE before you maintain the processing module.
Refer to
 General
Before you start this procedure, ensure you are logged in to the processing module. Refer to
in and log
out.
BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
Slide type
IHC slide
ISH slide
Multiplex 2 to 6
cautions.
1.
Equivalent slide count
1
2
2
Tap Status.
Chapter  4 Cleaning and maintenance
2.1 Log