5.4
Remove a slide fragment from an ARC Module
If you notice a slide with a portion missing in the Unload Drawer, you need to find and remove the slide fragment from
the ARC Module.
Do not load any new slides.
BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
5.
Close the Hood.
1.
Check the Action Queue to see if the error has been flagged
by the processing module. If not, then the slide breakage
has not been detected.
2.
Open the Preload Drawer and remove the slides, then wait
until all slides have finished processing.
Chapter 5 Troubleshooting
198