5.3.2
Manually retrieve slides from the Work Surface
WARNING: You must wear the minimum required PPE before you maintain the processing module.
Refer to
General
Before you start this procedure, ensure you are logged in to the processing module. Refer to
in and log
out.
BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
1.
Open the Preload and Unload Drawers and remove the
slides.
Check the drawers for debris. If there is debris, clean the
drawers. Refer to
Waste Drains and Sumps, and Pickup
2.
Close the Slide Preload and Unload Drawers.
cautions.
3.
Open the Preload and Unload Drawers and retrieve the
slides.
Check the drawers for debris. If there is debris, clean the
drawers. Refer to
Waste Drains and Sumps, and Pickup
Chapter 5 Troubleshooting
4.12 Clean the Slide Drawer Inserts,
Filter.
4.12 Clean the Slide Drawer Inserts,
Filter.
2.1 Log