Manually Retrieve Slides From Arc Modules During A Power Failure - Leica BOND-PRIME User Manual

Full y au toma ted ihc and i sh s taining s y s tem
Table of Contents

Advertisement

In the event of a power failure, refer to
power
failure.
5.3.4
Manually retrieve slides from ARC Modules during
a power failure
In the event of a power failure, you can follow these steps to manually remove slides.
WARNING: You must wear the minimum required PPE before you maintain the processing module.
Refer to
 General
BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
8.
Restart the processing module. Refer to
processing
5.3.4 Manually retrieve slides from ARC Modules during a
cautions.
1.
Power off the processing module.
module.
a.
Open the Reservoir Cabinet door.
b.
Turn the AC power switch counterclockwise.
c.
Close the Reservoir Cabinet door.
WARNING: Take care not to trip over the
Reservoir Cabinet door when it is in the open
position.
Chapter  5 Troubleshooting
3.2 Start the
196