Cleaning and Maintenance

6.1

Cleaning

• Before beginning to clean the RM CoolClamp, make sure that the power pack has been unplugged
and all surfaces have cooled down to below +50 °C (+122 °F).
• Clean the RM CoolClamp using paraffin remover only. Xylene and other solvents are not permitted.
Wait to reactivate it until all the liquid has evaporated.
• Make sure that no liquid ingresses into the instrument (ventilation slots). Use only enough liquid as is
absolutely necessary. The assembly group is not suited for cleaning by dipping into liquid.
• To guarantee that the RM CoolClamp functions flawlessly, make sure that the fan rotates and is not
blocked. In addition, make sure that the ventilation openings are free before and during use. refer to
(→ P. 9 – 4.2.1 Ventilation
• No cleaning in the incubator is permitted.
• The RM CoolClamp must not be disassembled for cleaning.
6.2

Maintenance

RM CoolClamp is maintenance free.

6.2.1 Malfunctions

• Report any malfunctions that occur to the person responsible for the instrument immediately.
• Make sure that the unit has been secured against misuse and abuse.
• Any safety-related parts that have been removed must be reattached and checked prior to instrument
setup.

6.2.2 Servicing the RM CoolClamp

The unit is not intended to be serviced. Servicing is accomplished by replacement. No repairs may be
carried out by the user.

6.2.3 Disposing of the unit

Dispose of the unit in accordance with the respective environmental directives of the respective
country.
In the event of contamination, comply with the safety instructions.
The RM CoolClamp is RoHS-compliant.