Bellows assembly tests

w WARNING
1009-0633-000
5. Attach the bottom edge of the bellows to the rim. Verify only the bottom
ring of the bellows is fitted over the rim.
6. Lower the housing and turn it clockwise to lock. Verify you cannot lift off of
the housing.
7. In the above steps, if you see a dust-like powder on the bellows housing
or on the bellows, apply a thin layer of KRYTOX lubricant to the ribs of the
bellows housing. Make sure the lubricant is applied smoothly and there
are no lumps.
8. Perform the Bellows assembly tests before completing the assembly of
the breathing system.
Objects in the breathing system can stop gas flow to the patient.
This can cause injury or death:
• Do not use a test plug that is small enough to fall into the
breathing system.
• Make sure that there are no test plugs or other objects caught
in the breathing system.
w
The bellows assembly test does not replace the preoperative
tests. Always complete the tests in the section Preoperative tests
before you use the system with a patient.
2 Cleaning and Sterilization
2-15