Table 2-3
Table 2-4
100 nA
100 μA
100 mA
Introduction
Measurement Modules
HPSMU Measurement Voltage Values and Resolutions
Measurement
Range
Value
0 ≤ |V| ≤ 2.2 V
2 V
0 ≤ |V| ≤ 22 V
20 V
0 ≤ |V| ≤ 44 V
40 V
0 ≤ |V| ≤ 110 V
100 V
0 ≤ |V| ≤ 200 V
200 V
HPSMU Measurement Current Values and Resolutions
Range
Measurement Value
0 ≤ |I| ≤ 1.15 nA
1 nA
0 ≤ |I| ≤ 11.5 nA
10 nA
0 ≤ |I| ≤ 115 nA
1 μA
0 ≤ |I| ≤ 1.15 μA
10 μA
0 ≤ |I| ≤ 11.5 μA
0 ≤ |I| ≤ 115 μA
0 ≤ |I| ≤ 1.15 mA
1 mA
0 ≤ |I| ≤ 11.5 mA
10 mA
0 ≤ |I| ≤ 115 mA
0 ≤ |I| ≤ 1 A
1 A
a. This column is applied to the auto ranging or the limited auto ranging. For
fixed ranging, maximum measurement value is Range column value.
2-12
Measurement Resolutions
a
High Speed ADC
100 μV
1 mV
2 mV
5 mV
10 mV
Measurement Resolutions
a
High Speed ADC
50 fA
500 fA
5 pA
50 pA
500 pA
5 nA
50 nA
500 nA
5 μA
50 μA
Keysight E5260/E5270 User's Guide, Edition 6
High Resolution ADC
2 μV
20 μV
40 μV
100 μV
200 μV
High Resolution ADC
10 fA
10 fA
100 fA
1 pA
10 pA
100 pA
1 nA
10 nA
100 nA
1 μA