Overhead Time; To Specify Source Output Time - Keysight Technologies E5260 Series User Manual

Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

Function Details
Measurement Time

Overhead Time

The overhead time is the time required for range changing and so on. This time
depends on the measurement condition, and cannot be specified. Major elements of
the overhead time are:
Range changing time during measurement (when measurement ranging mode is
set to auto or limited auto)
Range changing time at measurement start (when the measurement range is less
than the compliance value)

To Specify Source Output Time

As shown in the discussion of the overhead time, you cannot specify the
measurement time. However, in sweep measurement mode, you can specify a
source output time that includes the measurement time. To specify the source output
time, set the delay time and the step delay time. The step delay time is defined as the
time from the measurement trigger up to the trigger for changing the source output
value. See the figure below.
Source output time
Step delay time (Tsd)
Delay time (Td)
Vstart
Hold time
The step delay time must cover the measurement time to keep the source output time
constant.
These values can be set by the WT command.
6-18
Td
Tsd
Measurement time
Keysight E5260/E5270 User's Guide, Edition 6
Td
Tsd

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5270 seriesE5260aE5262aE5263aE5270b

Table of Contents