CAL
NOTE
Displays the sub menu that provides the following functions.
EXECUTE
Displays the following message.
Calibration: ALL
Press the arrow key to select the slot to calibrate. Then, press the
start self-calibration, or press the
performing self-calibration.
During self-calibration, the following message will be displayed.
Press [Exit] to abort
Then, press the
Exit
completed and the following message is displayed:
Calibration: Pass or Fail
Press the
key to return to the previous menu.
Exit
RESULT
Displays the following message.
FRAME: Pass or Fail Error No.
Press the arrow key to display the result of another slot as shown below:
Slot#: Pass or Fail Error No.
where # is 1 or 2 for the 2-ch mainframe, 1, 2, ... or 8 for the 8-ch mainframe,
and Error No. is the error number.
Press the
key to return to the previous menu.
Exit
To execute CAL of Keysight E5270B installed with ASU
If you execute the CAL of the E5270B installed with the ASU (Atto Sense and
Switch Unit), the E5270B executes the self-calibration and the 1 pA range offset
measurement for the measurement channels connected to the ASUs. The offset data
is temporarily memorized until the E5270B is turned off, and is used for the
compensation of the data measured by the 1 pA range of the channels. The E5270B
performs the data compensation automatically and returns the compensated data.
Since the E5270B is turned on, if you do not execute the CAL, the E5270B performs
the data compensation by using the pre-stored offset data.
Keysight E5260/E5270 User's Guide, Edition 6
key to return to the previous menu without
Exit
key to abort self-calibration, or wait until self-calibration is
Front Panel Reference
Setup Menus
key to
Enter
4-27