170
Incorrect usage of internal variable.
The internal variable must be %In for integer data, or %Rn for real
data. where n is an integer, 0 to 99. Use %In for the integer type
command parameters; and use %Rn for the real type command
parameters. For the internal variables, refer to "VAR" command in
Programming Guide.
171
Internal variable is not allowed.
The internal variables %In and %Rn are not available for the ACH,
VAR, and VAR? commands. Do not use the internal variables for the
commands.
200
Channel output switch must be ON.
To enter the specified command, set the channel output switch to ON.
201
Compliance must be set.
To change the source output mode (voltage or current), set the
compliance value.
202
Interlock circuit must be closed.
To set the output voltage or the voltage compliance to more than ±42 V
(high voltage state), close the interlock circuit. If the interlock circuit is
opened in the high voltage state, outputs of all units will be set to 0 V.
203
Cannot enable channel.
The channel output switch cannot be set to ON in the high voltage state.
Set the output voltage or the voltage compliance to ±42 V or less to set
the switch to ON.
204
Cannot disable channel.
The channel output switch cannot be set to OFF in the high voltage
state. Set the output voltage or the voltage compliance to ±42 V or less
to set the switch to OFF. Or send the CL command with no parameter to
set switches of all channels to OFF immediately.
205
DZ must be sent before RZ.
The RZ command is effective for the channels set to 0 V output by the
DZ command.
Keysight E5260/E5270 User's Guide, Edition 6
If You Have a Problem
Error Codes
7-13