Keysight Technologies E5260 Series User Manual page 45

Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

Table 2-7
Table 2-8
MPSMU Measurement Voltage Values and Resolutions
Measurement
Range
Value
0 ≤ |V| ≤ 0.55 V
0.5 V
0 ≤ |V| ≤ 2.2 V
2 V
0 ≤ |V| ≤ 5.5 V
5 V
0 ≤ |V| ≤ 22 V
20 V
0 ≤ |V| ≤ 44 V
40 V
0 ≤ |V| ≤ 100 V
100 V
MPSMU Measurement Current Values and Resolutions
Measurement
Range
Value
0 ≤ |I| ≤ 1.15 nA
1 nA
0 ≤ |I| ≤ 11.5 nA
10 nA
0 ≤ |I| ≤ 115 nA
100 nA
1 μA
0 ≤ |I| ≤ 1.15 μA
10 μA
0 ≤ |I| ≤ 11.5 μA
100 μA
0 ≤ |I| ≤ 115 μA
0 ≤ |I| ≤ 1.15 mA
1 mA
0 ≤ |I| ≤ 11.5 mA
10 mA
0 ≤ |I| ≤ 100 mA
100 mA
a. This column is applied to the auto ranging or the limited auto ranging. For
fixed ranging, maximum measurement value is Range column value.
Keysight E5260/E5270 User's Guide, Edition 6
Measurement Resolutions
a
High Speed ADC
25 μV
100 μV
250 μV
1 mV
2 mV
5 mV
Measurement Resolutions
a
High Speed ADC
50 fA
500 fA
5 pA
50 pA
500 pA
5 nA
50 nA
500 nA
5 μA
Introduction
Measurement Modules
High Resolution ADC
0.5 μV
2 μV
5 μV
20 μV
40 μV
100 μV
High Resolution ADC
10 fA
10 fA
100 fA
1 pA
10 pA
100 pA
1 nA
10 nA
100 nA
2-15

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5270 seriesE5260aE5262aE5263aE5270b

Table of Contents