Keysight Technologies E5260 Series User Manual page 119

Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

AUX On/Off
(Shift+CPL Value)
OutRange
On/Off
SerRes
(Shift+On/Off)
The Keysight E5260 series does not have this function. This function is available for
the Keysight E5270B installed with the high resolution SMU (HRSMU).
To use the ASU (Atto Sense and Switch Unit), connect the ASU to the HRSMU
properly. The ASU adds the 1 pA range to the HRSMU. Connection must be
performed before the Keysight E5270B is turned on.
If you use other instrument such as C meter, connect the instrument to the AUX
input of the ASU. The ASU provides the input selection function.
This shift key operation is available when you are setting the HRSMU connected to
the ASU, and switches the ASU input resource (HRSMU or the instrument
connected to the AUX input) to be connected to the ASU output. When the E5270B
is turned on, the SMU side is connected to the ASU output.
When the AUX side is connected to the ASU output, the HRSMU cannot be used.
Then the Output indicator displays the reversed triangle mark.
The ASU is equipped with two LEDs that show which input is internally connected
to the output.
Displays the output range setup menu in the Change Setup area. See
Setup Menu" on page
4-32.
Enables or disables the source output of the channel specified by the S-ch status
indicator. Then the Output status indicator shows the source output ON or OFF
status.
Displays the following message.
[S-ch #] Series Resistor: ON or OFF
where # is 1 or 2 for the 2-ch mainframe, 1, 2, ... or 8 for the 8-ch mainframe.
Press the
key to specify the source channel, and press the arrow key to
OutCh
connect the series resistor (ON) to the channel or to disconnect it (OFF). Then press
the
key to make the setup effective, or press the
Enter
setup.
A series resistor (1 MΩ) is mounted on each module. If you use a series resistor, the
voltage you set is applied to the near side of the series resistor. Consequently, the
voltage will be divided by the series resistor and the device under test.
The series resistor cannot be used for the measurements that use the high resolution
SMU (HRSMU) connected to the Atto Sense and Switch Unit (ASU) or the
measurements that use 1 A range of the high power SMU (HPSMU).
Keysight E5260/E5270 User's Guide, Edition 6
Front Panel Reference
Front Panel Key Details
"Output Range
key to cancel changing the
Exit
4-15