WARNING
Installation
Maintenance

Self-test and Diagnostics

Keysight E5260/E5270 provides the following functions to check the operation.
Perform the following functions as necessary.
Self-test
Refer to
"To Perform Self-test" on page
Self-calibration
Refer to
"To Perform Self-Calibration" on page
Diagnostics
Refer to
"To Perform Diagnostics" on page
Interlock Open/Close test and High Voltage LED test
Interlock function limits the maximum output voltage to ± 42 V when the
measurement terminal is touchable. For safety, this function must be checked
before using the E5260/E5270 after power on at least once a day by performing
the diagnosis.
Test d'ouverture/de fermeture Interlock et test High Voltage LED
La fonction de verrouillage limite la tension de sortie maximale à ± 42 V
lorsque la borne de mesure peut être touchée. Pour des raisons de sécurité,
cette fonction doit être vérifiée avant d'utiliser le E5260/E5270 après la mise
sous tension, au moins une fois par jour en effectuant le diagnostic.
3-44
5-7.
5-8.
5-9.
Keysight E5260/E5270 User's Guide, Edition 6