Table 2-13
Table 2-14
HPSMU Output Voltage Ranges and Resolutions
Range
Output Value
0 ≤ |V| ≤ 2 V
2 V
0 ≤ |V| ≤ 20 V
20 V
0 ≤ |V| ≤ 40 V
40 V
0 ≤ |V| ≤ 100 V
100 V
0 ≤ |V| ≤ 200 V
200 V
a. 100 μV for the GPIB remote mode.
HPSMU Output Current Ranges and Resolutions
Range
Output Value
0 ≤ |I| ≤ 115 nA
100 nA
1 μA
0 ≤ |I| ≤ 1.15 μA
10 μA
0 ≤ |I| ≤ 11.5 μA
100 μA
0 ≤ |I| ≤ 115 μA
0 ≤ |I| ≤ 1.15 mA
1 mA
0 ≤ |I| ≤ 11.5 mA
10 mA
0 ≤ |I| ≤ 50 mA
100 mA
50 mA < |I| ≤ 115 mA
0 ≤ |I| ≤ 50 mA
1 A
50 mA < |I| ≤ 125 mA
125 mA < |I| ≤ 500 mA
500 mA < |I| ≤ 1 A
Keysight E5260/E5270 User's Guide, Edition 6
Measurement Modules
Setting
Maximum
Resolution
Current
a
± 1000 mA
1 mV
1 mV
± 1000 mA
2 mV
± 500 mA
5 mV
± 125 mA
10 mV
± 50 mA
Setting
Maximum
Resolution
Voltage
5 pA
± 200 V
50 pA
± 200 V
500 pA
± 200 V
5 nA
± 200 V
50 nA
± 200 V
500 nA
± 200 V
5 μA
± 200 V
5 μA
± 100 V
50 μA
± 200 V
50 μA
± 100 V
50 μA
± 40 V
50 μA
± 20 V