Table 2-19
Table 2-20
MPSMU Measurement Voltage Values and Resolutions
Range
2 V
20 V
40 V
100 V
MPSMU Measurement Current Values and Resolutions
Range
100 nA
1 μA
10 μA
100 μA
1 mA
10 mA
100 mA
200 mA
a. This column is applied to the auto ranging or the limited auto ranging.
For fixed ranging, maximum measurement value is Range column
value.
Keysight E5260/E5270 User's Guide, Edition 6
a
Measurement Value
0 ≤ |V| ≤ 2.2 V
0 ≤ |V| ≤ 22 V
0 ≤ |V| ≤ 44 V
0 ≤ |V| ≤ 100 V
a
Measurement Value
0 ≤ |I| ≤ 115 nA
0 ≤ |I| ≤ 1.15 μA
0 ≤ |I| ≤ 11.5 μA
0 ≤ |I| ≤ 115 μA
0 ≤ |I| ≤ 1.15 mA
0 ≤ |I| ≤ 11.5 mA
0 ≤ |I| ≤ 115 mA
0 ≤ |I| ≤ 200 mA
Introduction
Measurement Modules
Measurement Resolutions
100 μV
1 mV
2 mV
5 mV
Measurement Resolutions
5 pA
50 pA
500 pA
5 nA
50 nA
500 nA
5 μA
10 μA
2-25