Error Codes; Operation Error - Keysight Technologies E5260 Series User Manual

Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

Error Codes

If errors occur, error codes are stored in the error buffer. To read the error code,
execute the "ERR?" command. To read the error message, execute the "EMG?"
command.
The output of the error codes is in the order that they occurred, and the first four
error codes are stored in the buffer. If no errors occurred, "0, 0, 0, 0" is returned.

Operation Error

100
Undefined GPIB command.
Send the correct command.
102
Incorrect numeric data syntax.
Correct the data syntax.
103
Incorrect terminator position.
Correct the command syntax. The number of parameters will be
incorrect.
120
Incorrect parameter value.
Correct the parameter value.
121
Channel number must be 1 to 2, or 1 to 8.
Correct the channel number. The channel number must be 1 to 2 for the
Keysight E5262A/E5263A, or 1 to 8 for the Keysight
E5260A/E5270B.
122
Number of channels must be corrected.
Check the MM, FL, CN, CL, IN, DZ, or RZ command, and correct the
number of channels.
123
Compliance must be set correctly.
Incorrect compliance value was set. Set the compliance value correctly.
124
Incorrect range value for this channel.
Check the range value available for the channel, and correct the range
value.
Keysight E5260/E5270 User's Guide, Edition 6