To Perform Self-Calibration - Keysight Technologies E5260 Series User Manual

Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

NOTE
NOTE
Front Panel Operations
Setting Up Keysight E5260/E5270

To Perform Self-Calibration

1. Open the measurement terminals of the Keysight E5260/E5270.
2. Press the
key and the
Shift
3. Move the cursor to CAL, then press the
4. Move the cursor to EXECUTE, then press the
5. Press the arrow keys to select the slot to calibrate; ALL or Slot n (n: 1 or 2 for
2-ch mainframe, 1 to 8 for 8-ch mainframe).
6. Press the
key to start calibration, or press the
Enter
During calibration, press the
7. Press the
key three times to close the setup menu.
Exit
To display calibration results
1. Press the
key and the
Shift
2. Move the cursor to CAL, then press the
3. Move the cursor to RESULT, then press the
4. Press the arrow keys to select the calibration item; FRAME or Slot n (n: 1 or 2
for 2-ch mainframe, 1 to 8 for 8-ch mainframe). The calibration result is
displayed for each calibration item.
5. Press the
key three times to close the setup menu.
Exit
To use a module that failed self-calibration, refer to
self-test" on page
5-7. The procedure also applies to modules that failed
self-calibration.
If you execute the CAL of the E5270B installed with the ASU (Atto Sense and
Switch Unit), the E5270B executes the self-calibration and the 1 pA range offset
measurement for the measurement channels connected to the ASUs. The offset data
is temporarily memorized until the E5270B is turned off, and is used for the
compensation of the data measured by the 1 pA range of the channels. The E5270B
performs the data compensation automatically and returns the compensated data.
Since the E5270B is turned on, if you do not execute the CAL, the E5270B performs
the data compensation by using the pre-stored offset data.
5-8
key.
Menu
key.
Enter
Enter
key to abort it.
Exit
key.
Menu
key.
Enter
Enter
"To use modules that failed
Keysight E5260/E5270 User's Guide, Edition 6
key.
key to cancel calibration.
Exit
key.

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5270 seriesE5260aE5262aE5263aE5270b

Table of Contents