Configuration; Moving Around In The Menus - Rice Lake IQ plus 800 Installation & Operation Manual

Digital weight indicator
Table of Contents

Advertisement

3.

Configuration

NOTE:
®
HOTLINE
is a PC program
that uses serial communication
to configure IQ plus 800/810
indicators running standard
indicator software.
See your distributor for more
information.
Warning
!
Established digital outputs
may still be activated when
configuring, calibrating, or
changing setpoint param-
eters. De-energize any
external equipment con-
trolled by the indicator's
digital outputs while configu-
ration, calibration, or
setpoint changes are being
made.
Caution
!
When exploring the various
menus, be careful not to
accidentally change the
parameter settings. Remem-
ber, the display shows the
default setting when you first
move to a particular level; so
make sure the same setting
appears on the display
before you press DISP TARE
to exit.
To access the setup mode, open the case and move switch SW1 on the CPU
board (see Figure 2-3 in Section 2, Installation, for location). This 2-position
switch is near the middle of the board, at the edge. When you move it to the
setup position, "CONFIG" appears on the display. "CONFIG" is one of
eleven main menu items, each with its own set of configurable parameters.

MOVING AROUND IN THE MENUS

The parameter menus (shown below) are arranged under eleven main group
headings, with various levels under each heading. See Figures 3-1 through
3-10 for menu charts of each parameter selection.
To set up the unit, you navigate through the configuration menus with four
front-panel keys (shown below) that become directional keys in the setup
mode. Press SET POINT and CLEAR to scroll left and right (horizontally)
on the same level. Press DISP TARE and TIME/DATE to move up and
down (vertically) to different menu levels.
To select a parameter, press SET POINT or CLEAR until the desired main
menu group appears on the display. Then, press TIME/DATE to move
down to the desired level. When moving down through the menus, the
default setting appears first on the display. To change a default, scroll left or
right through the various options for that level. When the desired option
appears on the display, press DISP TARE to lock in your selection and
move back up one level. For parameters requiring a numerical entry, key in
the number, press ENTER, and scroll up to lock in the number.
Move up to
previous menu
level.
Move down to
next menu
level.
DISP
SET
TARE
POINT
TIME/
CLEAR
DATE
Move left on
same menu
level.
Move right on
same menu level.
Configuration